Z-1127-05
Depuy LCS Complete RP insert; polyethylene tibial insert STD+, 12.5 mm; Ref 1294-05-512
Depuy Orthopaedics Inc. / initiated 2005-05-26 / Terminated / root cause: packaging
Depuy Orthopaedics, Inc. began a recall of Depuy LCS Complete RP insert; polyethylene tibial insert STD+, 12.5 mm; Ref 1294-05-512 on . The reason FDA records is "An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1127-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- Depuy LCS Complete RP insert; polyethylene tibial insert STD+, 12.5 mm; Ref 1294-05-512.
- Product code
- NJL Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Reason for recall
- An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
- Root cause tag
- packaging
- FDA root cause
- Other
- Quantity in commerce
- 169
- Distribution
- Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
- Product codes and lots
- Lots X2XB54, X2XBY4, X4TC34, X4TC94, X6AA64, X6AA84, X6AAW4, XW8C64, Y2RBX4, Y4AFY4, Y4AGE4, Y4AGF4, Y4AGG4, Y4HP24, Y4HPX4, YD2ES4 and YP2HN4.