Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1164-05

Depuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4,10 mm; Ref 1581-14-010

Depuy Orthopaedics Inc. / initiated 2005-05-26 / Terminated / root cause: packaging

Depuy Orthopaedics, Inc. began a recall of Depuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4,10 mm; Ref 1581-14-010 on . The reason FDA records is "An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1164-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
Depuy P.F.C. SIGMA curved insert; polyethylene tibial insert, Sz 4,10 mm; Ref 1581-14-010.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root cause tag
packaging
FDA root cause
Other
Quantity in commerce
12
Distribution
Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
Product codes and lots
Lot Y4PCW4.