Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0924-05

Cancellous Bone Screw, 6.5 x 50 mm, Model 0214807

Plus Orthopedics / initiated 2005-06-03 / Terminated / root cause: labeling and UDI

Plus Orthopedics USA began a recall of Cancellous Bone Screw, 6.5 x 50 mm, Model 0214807 on . The reason FDA records is "Screw is mislabeled as a 50 mm screw when in fact it is 45 mm long". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0924-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Plus Orthopedics
Product
Cancellous Bone Screw, 6.5 x 50 mm, Model 0214807
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Screw is mislabeled as a 50 mm screw when in fact it is 45 mm long.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Notification was made by telephone beginning June 2, 2005. Recall is on-going.
Quantity in commerce
54 USA 78 Europe
Distribution
Idaho, Texas and Virginia
Product codes and lots
Lot / Unit Code: 044152