Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0907-05

CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds)...

Boston Scientific Corporation / initiated 2005-06-17 / Terminated

Guidant Corporation began a recall of CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant... on . The reason FDA records is "Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0907-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
FDA root cause
Other
Action
Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.
Quantity in commerce
16,000 (Approximately 11,900 remain implanted worldwide)
Distribution
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom and Venezuela.
Product codes and lots
RENEWAL (model H135) serial number range 100027 - 999983 includes CRT-Ds manufactured on or before August 26, 2004. These ranges include all affected devices but may include some unaffected devices.