VENTAK PRIZM AVT model no. 1900, Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide...
Boston Scientific CRM Corp began a recall of VENTAK PRIZM AVT model no. 1900, Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100... on . The reason FDA records is "The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0954-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- VENTAK PRIZM AVT model no. 1900, Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Product code
- LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
- Reason for recall
- The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.
- FDA root cause
- Other
- Action
- Firm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.
- Quantity in commerce
- 937 devices (714 currently implanted worldwide)
- Distribution
- worldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Syrian Arab Republic, Turkey, US - Virgin Islands and United Kingdom.
- Product codes and lots
- all serial numbers