VITALITY AVT, model A135 and model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular...
Boston Scientific CRM Corp began a recall of VITALITY AVT, model A135 and model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide... on . The reason FDA records is "The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0955-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- VITALITY AVT, model A135 and model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Product code
- LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
- Reason for recall
- The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.
- FDA root cause
- Other
- Action
- Firm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.
- Quantity in commerce
- 22,575 devices (19,446 implanted world wide)
- Distribution
- worldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Syrian Arab Republic, Turkey, US - Virgin Islands and United Kingdom.
- Product codes and lots
- all serial numbers