Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0966-05

Boston Scientific brand Standard PEG (percutaneous endoscopic gastrostomy tubes) Kit, 24 Fr, push; catalog...

Boston Scientific Corporation / initiated 2005-06-02 / Terminated

Boston Scientific Corp began a recall of Boston Scientific brand Standard PEG (percutaneous endoscopic gastrostomy tubes) Kit, 24 Fr, push; catalog number 6825 on . The reason FDA records is "Two blades may be heat staked onto the stiletto instead of the expected one blade". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0966-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific brand Standard PEG (percutaneous endoscopic gastrostomy tubes) Kit, 24 Fr, push; catalog number 6825.
Product code
KGC Tube, Gastro-Enterostomy
Reason for recall
Two blades may be heat staked onto the stiletto instead of the expected one blade.
FDA root cause
Other
Action
Customers were notified by a recall letter dated 6/7/05 which instructed them to return any of the recalled products on hand.
Distribution
Nationwide and Panama.
Product codes and lots
Lot 7404367.