Z-0970-05

Boston Scientific brand One Step Button (percutaneous endoscopic gastrostomy tubes) Kit, 24/2.4 pull; catalog...

Boston Scientific Corporation / initiated 2005-06-02 / Terminated

Boston Scientific Corp began a recall of Boston Scientific brand One Step Button (percutaneous endoscopic gastrostomy tubes) Kit, 24/2.4 pull; catalog number 6851 on . The reason FDA records is "Two blades may be heat staked onto the stiletto instead of the expected one blade". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0970-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific brand One Step Button (percutaneous endoscopic gastrostomy tubes) Kit, 24/2.4 pull; catalog number 6851.
Product code
KGC Tube, Gastro-Enterostomy
Reason for recall
Two blades may be heat staked onto the stiletto instead of the expected one blade.
FDA root cause
Other
Action
Customers were notified by a recall letter dated 6/7/05 which instructed them to return any of the recalled products on hand.
Distribution
Nationwide and Panama.
Product codes and lots
Lots 7471765, 7471766, 7434813, 7471767 and 7434812.