CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and...
Guidant Corporation began a recall of CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and M179). Cardiac Resynchronization Therapy... on . The reason FDA records is "Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0952-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- CONTAK RENEWAL 4 AVT CRT-D (model no. M170 and M175) and CONTAK RENEWAL 4 AVT HE CRT-D (model no. M177 and M179). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, 4100 Hamiline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 AVT devices provide both atrial and ventricular tachyarrhythmia and cardiac resynchronization therapies.
- Product code
- NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Reason for recall
- Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.
- FDA root cause
- Other
- Action
- Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.
- Quantity in commerce
- 1423
- Distribution
- worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Denmark, Eygpt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Labanon, Luxembourg, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom and Venezuela.
- Product codes and lots
- all serial numers