Z-1045-05
Mammomat Novation Digital Mammography System. Model number 6646900
Siemens Medical Solutions USA, Inc. / initiated 2005-05-25 / Terminated
Siemens Medical Solutions USA, Inc began a recall of Mammomat Novation Digital Mammography System. Model number 6646900 on . The reason FDA records is "potential issue (image acquisition may become interrupted) when using multiple network nodes". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1045-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mammomat Novation Digital Mammography System. Model number 6646900
- Reason for recall
- potential issue (image acquisition may become interrupted) when using multiple network nodes
- FDA root cause
- Other
- Action
- The recalling firm has respresentatives/engineers hand carrying Customer Advisory Letters to their customers. The engineers are temporarily reconfiguring the ssytem to allow routing to only two network nodes in order to prevent this issue from occuring.
- Quantity in commerce
- 33 units
- Distribution
- Nationwide. The products were shipped to medical facilities in AL, CA, FL, GA, IL, LA, MO, NC, NJ, NV, NY, SC, TX, VA, and WV.
- Product codes and lots
- Serial numbers 1001, 1009, 1019, 1020, 1021, 1024, 1035, 1038, 1039, 1044, 1045, 1046, 1047, 1051, 1052, 1054, 1055, 1058, 1060, 1063, 1064, 1105, 1107, 1108, 1114, 1208, 1209, 1210, 1211, 1212, 1214, 1216, 1217, 1220, 1224, 1227, 1228, 1229, and 1234.