Z-0990-05
Axiom Aristos FX. Model number 74 14 803
Siemens Medical Solutions USA, Inc. / initiated 2005-06-10 / Terminated
Siemens Medical Solutions USA, Inc began a recall of Axiom Aristos FX. Model number 74 14 803 on . The reason FDA records is "potential issue (system may lock up) when repeating patient registrations". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0990-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Axiom Aristos FX. Model number 74 14 803
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- potential issue (system may lock up) when repeating patient registrations
- FDA root cause
- Other
- Action
- The recalling firm issed a Customer Advisory Letter via certified mail to their cutomers informing them of the problem and information for the workaround to prevent problems for occurring until the software is upgraded.
- Quantity in commerce
- 72 units
- Distribution
- Nationwide. The products were shipped to hospitals in AZ, CA, CO, FL, GA, IL, IN, KS, KY, ME, MI, MO, NC, NE, NJ, NY, OK, PA, TN, TX, UT, VA, WA, and WI.
- Product codes and lots
- Serial numbers 1001 through 1003, 1005 through 1047, and 1049 through 1247.