CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant...
Guidant Corporation began a recall of CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant corporation, Cardiac Rhythm Management, 4100 Hamline Avenue... on . The reason FDA records is "Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0908-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- CONTAK RENEWAL 2 (model H155) Heart Failure Device. Sterilized with gaseous ethylene oxide. Guidant corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
- Product code
- NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Reason for recall
- Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
- FDA root cause
- Other
- Action
- Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.
- Quantity in commerce
- 16,000 (Approximately 11,900 remain implanted worldwide) - both products combined
- Distribution
- worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom and Venezuela.
- Product codes and lots
- RENEWAL 2 (model H155) serial number range 173277 - 777182 includes CRT-Ds manufactured on or before August 26, 2004.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
