Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1095-05

ChemoSite Implantable venous access sytem with printed Polyurethane Catheter and 9 French Introducer Kit...

Medtronic, Inc. / initiated 2005-06-30 / Terminated

United States Surgical Corporation began a recall of ChemoSite Implantable venous access sytem with printed Polyurethane Catheter and 9 French Introducer Kit Catalog Number: 120045 on . The reason FDA records is "The tear away introducer sheath may have cracked hubs/handles and /or improper peeling of the sheath during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1095-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
ChemoSite Implantable venous access sytem with printed Polyurethane Catheter and 9 French Introducer Kit Catalog Number: 120045
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
The tear away introducer sheath may have cracked hubs/handles and /or improper peeling of the sheath during use
FDA root cause
Other
Action
United States Surgical notified customers by letter dated June 30, 2005 requesting return of the recalled product.
Quantity in commerce
2393 units
Distribution
Nationwide Foreign: Australia, Canada, Germany,Israel, Uruguay, Chile,
Product codes and lots
Lot Numbers: N4G22 N4G215 N4G436 N4J29 N4J212 N4K218 N4L258 N4M213