Z-1020-05
Adapter for ACMI Hysteroscope Catalog Number: 55035
Boston Scientific Corporation / initiated 2005-07-28 / Terminated
Boston Scientific Marlbourgh began a recall of Adapter for ACMI Hysteroscope Catalog Number: 55035 on . The reason FDA records is "The Hysteroscope Adapter used with the HTA System may disassemble and result in a fluid leak". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1020-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Adapter for ACMI Hysteroscope Catalog Number: 55035
- Product code
- MNB Device, Thermal Ablation, Endometrial
- Reason for recall
- The Hysteroscope Adapter used with the HTA System may disassemble and result in a fluid leak.
- FDA root cause
- Other
- Action
- Boston Scientifc notified accounts by letter dated June 28, 2005 via Federal Express. Accounts are requested to check invnetory and return. Any adapters out of box have no lot identification and can be tested per instructions
- Distribution
- CA, DE, FL, MA, MO, MS, OH, NE, NY, PA, UT, WI
- Product codes and lots
- Lot Number: 20410191