Z-1576-05
ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System
Diasorin Inc. / initiated 2004-03-09 / Terminated
Diasorin Inc. began a recall of ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System on . The reason FDA records is "An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1576-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- ELx800 microtiter plate reader software program written for use with ORTHO HCV Version 3.0 ELISA Test System.
- Product code
- MZO Assay, Enzyme Linked Immunosorbent, Hepatitis C Virus
- Reason for recall
- An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay.
- FDA root cause
- Other
- Action
- Dear Customer letters, dated 03/08/2004 were sent to customers informing them of the specification error and instructing users to validate their assays manually.
- Quantity in commerce
- 128
- Distribution
- AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OR, PA, TN, TX, UT, VA, WI & WV
- Product codes and lots
- version 12 and older software