Z-1376-05
UltraLite Vehicles powered by Rascal Motorized Vehicles
Electric Mobility / initiated 2005-05-05 / Terminated
Electric Mobility Corp began a recall of UltraLite Vehicles powered by Rascal Motorized Vehicles on . The reason FDA records is "The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1376-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Electric Mobility
- Product
- UltraLite Vehicles powered by Rascal Motorized Vehicles.
- Product code
- ITI Wheelchair, Powered
- Reason for recall
- The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities.
- FDA root cause
- Other
- Action
- The firm sent out letters to their customers on 4/20/2005. The firm's sales representatives were also notified on 4/20/2005.
- Quantity in commerce
- 788 units
- Distribution
- The firm sells the vehicles via direct sales to end customers and also to company owned inventory with sales representatives. The distribution is nationwide.
- Product codes and lots
- All units distributed under Model Number UL370.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.