Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1376-05

UltraLite Vehicles powered by Rascal Motorized Vehicles

Electric Mobility / initiated 2005-05-05 / Terminated

Electric Mobility Corp began a recall of UltraLite Vehicles powered by Rascal Motorized Vehicles on . The reason FDA records is "The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1376-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Electric Mobility
Product
UltraLite Vehicles powered by Rascal Motorized Vehicles.
Product code
ITI Wheelchair, Powered
Reason for recall
The engager set screw may become loose. This problem may cause the potentiometer to become off-center thus affecting the end users stopping capabilities.
FDA root cause
Other
Action
The firm sent out letters to their customers on 4/20/2005. The firm's sales representatives were also notified on 4/20/2005.
Quantity in commerce
788 units
Distribution
The firm sells the vehicles via direct sales to end customers and also to company owned inventory with sales representatives. The distribution is nationwide.
Product codes and lots
All units distributed under Model Number UL370.