VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . VIRTUS...

Boston Scientific Corporation / initiated 2005-07-18 / Terminated

Guidant Corporation began a recall of VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . VIRTUS PLUS II pacemakers are multiprogrammable... on . The reason FDA records is "A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1037-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
VIRTUS PLUS II (model nos. 1380 and 1480) pacemaker. Sterilized with gaseous ethylene oxide. Guidant . VIRTUS PLUS II pacemakers are multiprogrammable pacemakers from Guidant. The family consists of both dual-chamber and single-chamber models, offering adaptive-rate therapy and providing various levels of therapeutic and diagnostic functionality. Two sensors are available; these adapt the pacing rate to the patient's changing metabolic demand.
Product code
LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
Reason for recall
A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
FDA root cause
Other
Action
INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
Quantity in commerce
94
Distribution
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
Product codes and lots
Affected devices may be contained in the following serial number ranges.  Model 1380: serial numbers 100000 thru 100044.  Model 1480: Serial numbers 500005 thru 500072.