CONTAK TR (Model no. 1241) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac...
Guidant Corporation began a recall of CONTAK TR (Model no. 1241) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St... on . The reason FDA records is "A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1039-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- CONTAK TR (Model no. 1241) pacemaker. Sterilized with gaseous ethylene oxide. Guidant Corporation , Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. The Guidant CONTAK TR CHFD congestive heart failure devices are multiprogrammable dual-chamber devices that offer adaptive-rate congestive heart failure pacing therapy. CONTAK TR devices use an accelerometer that allows the pacing rate to adapt to the patient's changing metabolic demand as it responds to patient activity.
- Product code
- LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
- Reason for recall
- A hermetic sealing component utilized in the device may experience a gradual degradation, resulting in a higher than normal moisture content within the pacemeaker case late in the device's service life.
- FDA root cause
- Other
- Action
- INITIAL RECALL Urgent medical Device Safety Information & Corrective Action letter dated 07/18/05 was issued to physicians listing clinical Implications and patient management recommendations. EXPANDED RECALL Press was issued January 23, 2006. Urgent Medical Device Safety Information & Corrective Action letter, dated 01/21/06, was issued to physicians listing a summary of previous communication to physicians, advisory population update, and information relative to a second population of devices manufactured between October 19, 1998 and December 5, 2000. This letter gives the observed rate of occurrence, clinical implications, projected rate of occurrence and recommendations.
- Quantity in commerce
- INITIAL RECALL 164. EXPANDED RECALL 110.
- Distribution
- worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Costa Rica, Cyprus, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guadeloupe, Guyana, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Polynesia, Portugal, Reunion, Russia, Saudi Arabia, Singapore,Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan ROC, Turkey, United Kingdom, Uruguay and Venezuela. EXPANDED RECALL additional countries include: China, Croatia, Korea. Lithuania, Madagascar, Maldives, Malta, Monaco, Nepal and Thailand.
- Product codes and lots
- Affected devices may be contained in the following serial number ranges. INITIAL RECALL serial numbers 200128 thru 200479. EXPANDED RECALL serial numbers 200076 thru 200958