Z-1397-05

Alsius CoolGard 3000 Temperature Regulation System

Alsius Corp. / initiated 2005-07-25 / Terminated

Alsius Corporation began a recall of Alsius CoolGard 3000 Temperature Regulation System on . The reason FDA records is "Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1397-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alsius Corp.
Product
Alsius CoolGard 3000 Temperature Regulation System
Product code
NCX System, Hypothermia, Intravenous, Cooling
Reason for recall
Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
FDA root cause
Other
Action
Firm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.
Quantity in commerce
207
Distribution
Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,
Product codes and lots
All units manufactured before April 12, 2005.