Z-1397-05
Alsius CoolGard 3000 Temperature Regulation System
Alsius Corp. / initiated 2005-07-25 / Terminated
Alsius Corporation began a recall of Alsius CoolGard 3000 Temperature Regulation System on . The reason FDA records is "Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1397-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alsius Corp.
- Product
- Alsius CoolGard 3000 Temperature Regulation System
- Product code
- NCX System, Hypothermia, Intravenous, Cooling
- Reason for recall
- Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
- FDA root cause
- Other
- Action
- Firm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.
- Quantity in commerce
- 207
- Distribution
- Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,
- Product codes and lots
- All units manufactured before April 12, 2005.