Z-1460-05
VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc...
Ortho-Clinical Diagnostics, Inc. / initiated 2005-07-19 / Terminated
Ortho-Clinical Diagnostics began a recall of VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs... on . The reason FDA records is "Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1460-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests.
- Product code
- LCD Enzyme Immunoassay, Gentamicin
- Reason for recall
- Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
- FDA root cause
- Other
- Action
- Letters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots.
- Quantity in commerce
- 443 boxes
- Distribution
- The product was shipped nationwide and to Ortho-Clinical/J&J affiliates in Australia, Canada, Singapore, England, France, Germany, Italy and Spain.
- Product codes and lots
- 01-5846, Exp. 31 May 2006; 01-6036 Exp 31 October 2006; 01-6239 Exp 31 January 2007