Z-0012-06

EBI Trochanteric 11mm Long Nail

Ebi, LLC / initiated 2005-06-24 / Terminated / root cause: design

EBI, L.P. began a recall of EBI Trochanteric 11mm Long Nail on . The reason FDA records is "Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0012-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ebi, LLC
Product
EBI Trochanteric 11mm Long Nail.
Product code
JDS Nail, Fixation, Bone
Reason for recall
Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.
Root cause tag
design
FDA root cause
Other
Action
The firm sent out via e-mail notification to their field representatives on 6/27/2005 to identify, remove and return to EBI all affected product from their inventories. One direct customer, Dr. Maylack was sent a recall letter on 7/7/2005.
Quantity in commerce
102 nails
Distribution
There are 9 sales representatives that are instructed to remove and return the EBI nail from nationwide customers.
Product codes and lots
Catalog No./Lot Nos: 28226/164522; 28228/164523; 28230/175576; 28234/175583; 28238/175589; 28248/164527; 28324/164529; 28326/164531; 28328/164532; 28330/175614; 28332/170182; 28334/175621; 28338/175630; 28342/166415; 28344/164540; 28348/164545.