Z-0012-06
EBI Trochanteric 11mm Long Nail
Ebi, LLC / initiated 2005-06-24 / Terminated / root cause: design
EBI, L.P. began a recall of EBI Trochanteric 11mm Long Nail on . The reason FDA records is "Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0012-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ebi, LLC
- Product
- EBI Trochanteric 11mm Long Nail.
- Product code
- JDS Nail, Fixation, Bone
- Reason for recall
- Upon evaluation, the firm found that the fatigue strength of these nails was below the required value as specified in the design specification.
- Root cause tag
- design
- FDA root cause
- Other
- Action
- The firm sent out via e-mail notification to their field representatives on 6/27/2005 to identify, remove and return to EBI all affected product from their inventories. One direct customer, Dr. Maylack was sent a recall letter on 7/7/2005.
- Quantity in commerce
- 102 nails
- Distribution
- There are 9 sales representatives that are instructed to remove and return the EBI nail from nationwide customers.
- Product codes and lots
- Catalog No./Lot Nos: 28226/164522; 28228/164523; 28230/175576; 28234/175583; 28238/175589; 28248/164527; 28324/164529; 28326/164531; 28328/164532; 28330/175614; 28332/170182; 28334/175621; 28338/175630; 28342/166415; 28344/164540; 28348/164545.