Ultraview SL Capnography Module, Model 91517 A mainstream and sidestream gas analyzer designed to measure the...
Spacelabs Medical Incorporated began a recall of Ultraview SL Capnography Module, Model 91517 A mainstream and sidestream gas analyzer designed to measure the concentration of carbon dioxide in a gas mixture... on . The reason FDA records is "The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0517-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Ultraview SL Capnography Module, Model 91517 A mainstream and sidestream gas analyzer designed to measure the concentration of carbon dioxide in a gas mixture, to aid in determining the patient''s ventilatory, circulatory, and metabolic status.
- Product code
- CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- Reason for recall
- The transducer used to compensate for pressure changes was mounted backwards on the printed circuit board assembly. The module will not compensate for pressure changes and CO2 readings will be inaccurate at higher altitudes.
- FDA root cause
- Other
- Action
- On 7/20/05 Spacelabs called the consignees that received the units to notify them of the hazard and also the schedule a service call to repair the affected capnography modules. The firm will confirm that all consignees have been updated with the appropriate PCBA and will audit the device recall records for completion.
- Quantity in commerce
- 11 units
- Distribution
- 8 units shipped to France, 1 unit to Brussels, Belgium, and 2 units to a medical center in WY.
- Product codes and lots
- Model 91517 Lot BMP007559