Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1597-05

Alto VR (model 615) and Alto 2 VR(model 625) Implantable cardioverter Defibrillator. Ventricular...

Ela Medical, Inc. / initiated 2005-07-19 / Terminated / root cause: manufacturing process

ela Medical Llc began a recall of Alto VR (model 615) and Alto 2 VR(model 625) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Single chamber arrhythmia detection... on . The reason FDA records is "Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1597-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ela Medical, Inc.
Product
Alto VR (model 615) and Alto 2 VR(model 625) Implantable cardioverter Defibrillator. Ventricular antichycardia pacing. Single chamber arrhythmia detection. Antibradycardia pacing. ELA medical 98, rue Maurice Amoux, 92120 Montrouge- FRANCE. These ICDs are designed to recognize and treat slow or fast Ventricular Tachycardia or Ventricular Fibrillation by continuously monitoring ventricular activity to identify persistent ventricular arrhythmias and to deliver appropriate therapies.
Product code
MRM Defibrillator, Implantable, Dual-Chamber
Reason for recall
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
Root cause tag
manufacturing process
FDA root cause
Other
Action
A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.
Quantity in commerce
51 total devices (1-model 615 & 50-model 625)
Distribution
Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.
Product codes and lots
devices manufactured between April 2003 and August 2004. Model 615 serial number: 237XG065.   Model 625 serial numbers: 243XI009, 243XI060, 243XI075, 243XI089, 246XI019, 246XI031, 246XI076, 249XI044, 249XI045,  249XI093, 249XI106, 249XI123, 249XI134, 304XI069,  304XI070, 304XI087, 304XI095, 304XI117, 307XI016,  307XI040, 307XI041, 307XI050, 307XI053, 307XI093, 307XJ016, 312XI007, 312XI009, 312XI016, 312XI033, 312XI033, 312XI041, 312XI095, 312XI097, 312XI124, 402XI027, 402XI070, 404XI007, 404XI010, 404XI028, 404XI029, 404XI031, 404XI044, 409XI001, 409XI007,  409XI013, 409XI015, 409XI018, 413XI004, 414XI017 and 415XI025.