Z-1546-05

SeNova Test Strips for Testing Glucose in Whole Blood. Blood glucose test strips for use with the Senova...

Chdiagnostics, LLC / initiated 2005-03-14 / Terminated / root cause: manufacturing process

CHdiagnostics began a recall of SeNova Test Strips for Testing Glucose in Whole Blood. Blood glucose test strips for use with the Senova brand meter only. For in vitro diagnostics use. Do not... on . The reason FDA records is "A suspect material used in the manufacturer of the product has degraded over time and consequently may cause falsely high readings when testing your blood glucose level". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1546-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Chdiagnostics, LLC
Product
SeNova Test Strips for Testing Glucose in Whole Blood. Blood glucose test strips for use with the Senova brand meter only. For in vitro diagnostics use. Do not store strips outside vial. Discard 3 months after opening. Distributed by CHdiagnostics, Plymouth, MN 55447, USA. Made in U.K.
Product code
LFR Glucose Dehydrogenase, Glucose
Reason for recall
A suspect material used in the manufacturer of the product has degraded over time and consequently may cause falsely high readings when testing your blood glucose level.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Sales Representatives will visit their distribution accounts to ensure affected product is removed. Distributors are asked to follow up with their accounts to ensure that no product is left with their customers. Nursing home customers will be mailed letters detailing the recall and not to use the affected lots.
Quantity in commerce
43,138 vials
Distribution
AL, FL, IA, IL, IN, KS, MD, MN, MO, MS, ND, NY, OH, PA, SD, TN and WI. Distribution to Greece, Czech Republic and Sinapore is typically done through Cambridge Sensors Limited, United Kingdom.
Product codes and lots
Lot numbers: 407061, 4071601, 4071901, 4071902, 4072001, 4072002, 4072201, 4072202, 4072801, 4072802, 4072901, 4072902, 4073001; 4080301, 4080302, 4080303, 4080401, 4080402, 4080601, 4081101, 4081102, 4081601, 4081901, 4082001, 4082002, 4082301, 4082501, 4082502, 4082701, 4083101, 4083102, 4090101, 4090201, 4091601, 4091702, 4092101, 4092201, 4092301, 4092401, 4092701, 4092801, 4092901, 4100101, 4100401, 4100402, 4100501, 4100601, 4101101, 4101201, 4101301, 4101401, 4101801, 4102201, 4102502, 4102601, 4102701, 4102801, 4102901, 4110401, 4110501, 4111101, 4111601 and 4111602.