CHdiagnostics began a recall of SeNova Test Strips for Testing Glucose in Whole Blood. Blood glucose test strips for use with the Senova brand meter only. For in vitro diagnostics use. Do not... on . The reason FDA records is "A suspect material used in the manufacturer of the product has degraded over time and consequently may cause falsely high readings when testing your blood glucose level". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1546-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Chdiagnostics, LLC
- Product
- SeNova Test Strips for Testing Glucose in Whole Blood. Blood glucose test strips for use with the Senova brand meter only. For in vitro diagnostics use. Do not store strips outside vial. Discard 3 months after opening. Distributed by CHdiagnostics, Plymouth, MN 55447, USA. Made in U.K.
- Product code
- LFR Glucose Dehydrogenase, Glucose
- Reason for recall
- A suspect material used in the manufacturer of the product has degraded over time and consequently may cause falsely high readings when testing your blood glucose level.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Sales Representatives will visit their distribution accounts to ensure affected product is removed. Distributors are asked to follow up with their accounts to ensure that no product is left with their customers. Nursing home customers will be mailed letters detailing the recall and not to use the affected lots.
- Quantity in commerce
- 43,138 vials
- Distribution
- AL, FL, IA, IL, IN, KS, MD, MN, MO, MS, ND, NY, OH, PA, SD, TN and WI. Distribution to Greece, Czech Republic and Sinapore is typically done through Cambridge Sensors Limited, United Kingdom.
- Product codes and lots
- Lot numbers: 407061, 4071601, 4071901, 4071902, 4072001, 4072002, 4072201, 4072202, 4072801, 4072802, 4072901, 4072902, 4073001; 4080301, 4080302, 4080303, 4080401, 4080402, 4080601, 4081101, 4081102, 4081601, 4081901, 4082001, 4082002, 4082301, 4082501, 4082502, 4082701, 4083101, 4083102, 4090101, 4090201, 4091601, 4091702, 4092101, 4092201, 4092301, 4092401, 4092701, 4092801, 4092901, 4100101, 4100401, 4100402, 4100501, 4100601, 4101101, 4101201, 4101301, 4101401, 4101801, 4102201, 4102502, 4102601, 4102701, 4102801, 4102901, 4110401, 4110501, 4111101, 4111601 and 4111602.