Z-0624-06

In vitro Test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Reagent Pack 100 test TDx/TDxFLx Cyclosporine...

Abbott Diagnostic International, Ltd. / initiated 2005-03-23 / Terminated / root cause: manufacturing process

Abbott Health Products, Inc. began a recall of In vitro Test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Reagent Pack 100 test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Calibrators are composed of... on . The reason FDA records is "Patient results reported using calibrators made from new primary material may be falsely decreased by 5-8% compared to results using calibrators made from the previous primary material. An Important Product Information...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0624-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Diagnostic International, Ltd.
Product
In vitro Test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Reagent Pack 100 test TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Calibrators are composed of six vials of calibrators A, B, C, D, E, and F containing cyclosporine prepared in human blood component (non-reactive for HBxAg, anti-HCv, anti-HIV-1/HIV-20. TDx/TDxFLx Cyclosporine Monoclonal Whole Blood Controls are composed of three vials L, M and H containing cyclosporine in human blood componenet (nonreactive for HBsAg, anti-HCV, and anti-HIV-1/HIV-2).
Product code
MGU Fluorescence Polarization Immunoassay For Cyclosporine
Reason for recall
Patient results reported using calibrators made from new primary material may be falsely decreased by 5-8% compared to results using calibrators made from the previous primary material. An Important Product Information Letter with technical information regarding the Whole Blood Calibrators and Controls, implementing new internal reference standards was issued.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Abbott has recently implemented new internal reference standards for TDx/TDxFLx Cyclosporine monoclonal Whole Blood Calibrators and Controls. A change in performance is associated with the use of calibrator lots 1829Q100 and higher and control lots 18230Q100 and higher, which are the first lots manufactured using the new internal reference standards. The specific performance changes are identified in the letter, dated March 23, 2005, as compared to current product using the previous standards (calibrators lot 05434Q100 and control lot 03071Q100).
Quantity in commerce
4,514 kits were distributed
Distribution
Nationwide, Puerto Rico, Hawaii, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Chile, Uruguay, Argentina, Brazil, Dominican Republic, Canada, Germany, Japan , Singapore, India, Taiwan, Thailand, Korea, Australia, Honduras, and Panama,
Product codes and lots
Lot 18229Q100 (exp 8/26/2006) Lot 18230Q100 (exp 10/05/2006)