product labeled in part: ''***banicide ADVANCED, A Reusable Sterilizing & Disinfecting Solution***Active...
Pascal Co Inc began a recall of product labeled in part: ''***banicide ADVANCED, A Reusable Sterilizing & Disinfecting Solution***Active Ingredient: Glutaraldehyde (1,5 Pentanedial)... on . The reason FDA records is "The product, Banicide Advanced, a resuable sterilizing and disinfecting solution, is likely to be ineffective for its intended use because testing found it failed testing for sporicidal use". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1588-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Pascal International, Inc.
- Product
- product labeled in part: ''***banicide ADVANCED, A Reusable Sterilizing & Disinfecting Solution***Active Ingredient: Glutaraldehyde (1,5 Pentanedial) 3.5%***NET CONTENTS: 3.8l (ONE U.S. GALLON)***Manufactured by Pascal Company, Inc. 2929 N.E. Northup Way P.O. Box 1478 Bellevue, WA 98009-1478 USA***''
- Product code
- MED Sterilant, Medical Devices
- Reason for recall
- The product, Banicide Advanced, a resuable sterilizing and disinfecting solution, is likely to be ineffective for its intended use because testing found it failed testing for sporicidal use.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- On August 10, 2005 the firm began notifying consignees via phone and began follow up with notification letters on August 12, 2005. Consignees are being asked to return product. The firm subsequently extended its recall after lots manufactured in June 2005 were tested by FDA and failed the sporicidal testing. The firm identified recipients of the affected product and began to contact them by phone on October 3, 2005. The firm began follow-up notification via mail on October 6, 2005. All customers are requested to respond directly to Pascal regarding disposition of any remaining product.
- Quantity in commerce
- 89,922 units (each unit is one gallon)
- Distribution
- To dental product distributors nationwide and to 1 dental product distributor in Israel. Also directly to dental offices nationwide.
- Product codes and lots
- the following 3 codes were recalled initially: (LOT EXP 110305, LOT EXP 110605, LOT EXP 110705) On October 3, 2005 the firm expanded the recall to include all product produced between October 2004 through August 2005. Following are all the recalled lot codes. 100105, 100205, 100305, 100405, 100505, 100605, 110105, 110205, 110305, 110405, 110505, 110605, 110705, 120105, 120205, 120305, 010106, 010206, 010306, 010406, 010506, 020106, 020206, 020306, 020406, 020506, 020606, 020706, 020806, 030106, 030206, 030306, 030406, 030506, 030606, 030706, 030806, 030906, 031006, 031106, 040106, 040206, 040306, 050106, 050206, 050306, 050406, 060106, 060206, 060306, 060406, 060506, 070106, 070206, 070306, 070406, 080106,