Z-0338-06
Bivona brand Cuffless Extra Length Fixed Hyperflex Adult Tracheostomy Tube, 110 mm, sterile, Product code...
Abbott Medical / initiated 2005-04-19 / Terminated / root cause: manufacturing process
Smiths Medical ASD, Inc. began a recall of Bivona brand Cuffless Extra Length Fixed Hyperflex Adult Tracheostomy Tube, 110 mm, sterile, Product code 60AFHXL70 on . The reason FDA records is "The shaft was manufactured approximately 1 to 3 mm under the length specification, and may be 105 mm in length". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0338-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Bivona brand Cuffless Extra Length Fixed Hyperflex Adult Tracheostomy Tube, 110 mm, sterile, Product code 60AFHXL70.
- Product code
- BTO Tube, Tracheostomy (W/Wo Connector)
- Reason for recall
- The shaft was manufactured approximately 1 to 3 mm under the length specification, and may be 105 mm in length.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Consignee was notified via recall letter dated April 19, 2005.
- Quantity in commerce
- 11
- Distribution
- California, Indiana and North Carolina.
- Product codes and lots
- Lot G653628.
FIRST ARTICLE / WEEKLY