Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1457-05

Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP

Pioneer Surgical Technology began a recall of Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP on . The reason FDA records is "The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1457-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
Product
Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.
Product code
MVH Device, Detection, Sulfide
Reason for recall
The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.
FDA root cause
Other
Action
Distributors were notified via letter dated July 29, 2005 to retrieve all caps from inventory and point of use locations and to return them to Pioneer.
Quantity in commerce
5069
Distribution
Nationwide.
Product codes and lots
Lots 306456, 306457, 306613, 306712, 306713, 306714, 306816, 306817, 306818, 306819, 306820, 306821, 306822, 307526, 307527, 307528, 307667, 307668, 308510, 308938, 310662, 310663, 310664, 310665, 311718, 312169, 519935 and 521042.