Z-1400-05

Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100

Abbott Medical / initiated 2005-08-05 / Terminated / root cause: manufacturing process

Abbott Laboratories Diagnostic Div began a recall of Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100 on . The reason FDA records is "The product issue is that the value sheet for the Clinical Chemistry Bilirubin Calibrator, lot #20881M100 calibrator value sheet, has the calibrator values transposed for the Total Bilirubin Assay LN 7D60, Total...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1400-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
The product issue is that the value sheet for the Clinical Chemistry Bilirubin Calibrator, lot #20881M100 calibrator value sheet, has the calibrator values transposed for the Total Bilirubin Assay LN 7D60, Total Bilirubin Assay LN 8G62, and Neonatal Bilirubin Assay LN 9D88.
Root cause tag
manufacturing process
FDA root cause
Other
Action
A Device Correction letter correcting the transposition of the calibrator values dated August 5, 2005, was mailed to all customers receiving the Clinical Chemistry Bilirubin Calibrator, LN 1E66-03, Lot Number 20881M100.
Quantity in commerce
213
Distribution
Nationwide and Argentian, Brazil, Canada, Cayman, Chile, Columbia, Germany, Hong Kong, Jamaica, New Zealand, Puerto Rico, Singapore, Thailand, Venzuela
Product codes and lots
List # 1E66-03; Lot #20881M100