Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100
Abbott Laboratories Diagnostic Div began a recall of Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100 on . The reason FDA records is "The product issue is that the value sheet for the Clinical Chemistry Bilirubin Calibrator, lot #20881M100 calibrator value sheet, has the calibrator values transposed for the Total Bilirubin Assay LN 7D60, Total...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1400-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott Clinical Chemistry Bilirubin Calibrator, List # 1E66-03; Lot # 20881M100
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- The product issue is that the value sheet for the Clinical Chemistry Bilirubin Calibrator, lot #20881M100 calibrator value sheet, has the calibrator values transposed for the Total Bilirubin Assay LN 7D60, Total Bilirubin Assay LN 8G62, and Neonatal Bilirubin Assay LN 9D88.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- A Device Correction letter correcting the transposition of the calibrator values dated August 5, 2005, was mailed to all customers receiving the Clinical Chemistry Bilirubin Calibrator, LN 1E66-03, Lot Number 20881M100.
- Quantity in commerce
- 213
- Distribution
- Nationwide and Argentian, Brazil, Canada, Cayman, Chile, Columbia, Germany, Hong Kong, Jamaica, New Zealand, Puerto Rico, Singapore, Thailand, Venzuela
- Product codes and lots
- List # 1E66-03; Lot #20881M100