Product is a Pruitt Aortic Occlusion Catheter contained in a molded tray. The product is labeled in part...
LeMaitre Vascular, Inc. began a recall of Product is a Pruitt Aortic Occlusion Catheter contained in a molded tray. The product is labeled in part: ''***LeMaitre VASCULAR***Distributed By:LeMaitre... on . The reason FDA records is "LeMaitre Vascular is recalling certain lots of its Pruitt Aortic Occlusion Catheters due to a packaging defect which could compromise product sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0219-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LeMaitre Vascular Inc.
- Product
- Product is a Pruitt Aortic Occlusion Catheter contained in a molded tray. The product is labeled in part: ''***LeMaitre VASCULAR***Distributed By:LeMaitre Vascular, Inc.***Manufactured By: LeMaitre Vascular, Inc. 3101 37th Avenue North St. Petersburg, FL 33713***Pruitt Aortic Occlusion Catheter***REF 2101-12***STERILE***''.
- Product code
- DQT Occluder, Catheter Tip
- Reason for recall
- LeMaitre Vascular is recalling certain lots of its Pruitt Aortic Occlusion Catheters due to a packaging defect which could compromise product sterility.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- The distributors in Europe and Japan have been notified to contact their customers to return the products. All customers have been notified by letter.
- Quantity in commerce
- 1,919
- Distribution
- There are two lots in distribution (175 pieces) that may contain the small hole in the tray. These two lots were distributed in Japan and Europe. There are 16 lots in distribution (1,744 pieces) that utilize the trays with the thin spots that can potentially crack. These 16 lots were distributed in the United States, Canada, Japan and the European Union.
- Product codes and lots
- Lot No: 1010505-07, 1010505-08, 1030404-09, 104104-05, 1041404-06, 1042204-12, 1042704-01, 1042704-03, 1052204-01, 1061404-04, 1061404-05, 1070204-03, 1080204-08, 1082304-06, l100604-01, l100704-01, l120104-02, l121604-04.