Z-0131-06
Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH...
Tosoh Bioscience Inc / initiated 2005-08-18 / Terminated
Tosoh Bioscience, Inc. began a recall of Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH Bioscience Inc., 347 Oyster Point Blvd. Suite 201... on . The reason FDA records is "The firm has received reports that vials of the Level II were exhibiting poor and inconsistent chromatogram quality thus affecting control results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0131-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Tosoh Bioscience Inc
- Product
- Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH Bioscience Inc., 347 Oyster Point Blvd. Suite 201, South San Francisco, CA 94080.
- Product code
- JPK Mixture, Hematology Quality Control
- Reason for recall
- The firm has received reports that vials of the Level II were exhibiting poor and inconsistent chromatogram quality thus affecting control results.
- FDA root cause
- Other
- Action
- On August 18, 2005, the firm initiated the recall and issued notification to its consignees via mail, including instructions for distributors of the recall.
- Quantity in commerce
- 3,284 boxes
- Distribution
- The product was distributed to a total of 384 consignees throughout the US as well as Canada, Puerto Rico, Chile and Brazil.
- Product codes and lots
- Kit Lot Number: 31205; Controls I & II Lot Numbers: 31009AA, 31010AA