Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0131-06

Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH...

Tosoh Bioscience Inc / initiated 2005-08-18 / Terminated

Tosoh Bioscience, Inc. began a recall of Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH Bioscience Inc., 347 Oyster Point Blvd. Suite 201... on . The reason FDA records is "The firm has received reports that vials of the Level II were exhibiting poor and inconsistent chromatogram quality thus affecting control results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0131-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tosoh Bioscience Inc
Product
Tosoh brand Glycohemoglobin A1c Controls I & II, Catalog Number: 992133; Product is distributed by TOSOH Bioscience Inc., 347 Oyster Point Blvd. Suite 201, South San Francisco, CA 94080.
Product code
JPK Mixture, Hematology Quality Control
Reason for recall
The firm has received reports that vials of the Level II were exhibiting poor and inconsistent chromatogram quality thus affecting control results.
FDA root cause
Other
Action
On August 18, 2005, the firm initiated the recall and issued notification to its consignees via mail, including instructions for distributors of the recall.
Quantity in commerce
3,284 boxes
Distribution
The product was distributed to a total of 384 consignees throughout the US as well as Canada, Puerto Rico, Chile and Brazil.
Product codes and lots
Kit Lot Number: 31205; Controls I & II Lot Numbers: 31009AA, 31010AA