INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language...
Event Medical Ltd / initiated 2005-08-12 / Terminated
Event Medical began a recall of INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English... on . The reason FDA records is "Firm received reports of combustion within the blender system of the Inspiration Ventilator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0062-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Event Medical Ltd
- Product
- INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English (US); FR= French; DE=German; IT=Italian; PT=Portuguese; ES=Spanish; JP=Japanese; PL=Polish; RU=Russian
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Firm received reports of combustion within the blender system of the Inspiration Ventilator.
- FDA root cause
- Other
- Action
- Customer shall be notified by an Urgent Product Safety Notfication letter dated August 12, 2005. The letter shall advise the customer with the specifics of this problem, summarise the fact that the hazard potential in negligible and will advise the customer regarding options for replacement of the Blender Module assembly (part number F710400) and the Air/O2 Filters (part number F910300). The letter will also provide the customers with a set of actions which can be completed immediately in order to ensure the integrity of their gas supplies until their unit has been upgraded.
- Quantity in commerce
- 1,602 units
- Distribution
- Nationwide and Canada, Europe, Japan, Latin America, Asia/Pacific, Middle East
- Product codes and lots
- All serial numbers for Models F7100000-XX, F7200000-XX, and F7300000-XX