Z-0062-06

INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language...

Event Medical Ltd / initiated 2005-08-12 / Terminated

Event Medical began a recall of INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English... on . The reason FDA records is "Firm received reports of combustion within the blender system of the Inspiration Ventilator". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0062-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Event Medical Ltd
Product
INSPIRATION Ventilator Model Numbers: F7100000-XX, F7200000-XX, F7300000-XX; where XX designates the language variant. IN = English (International); GB=English (US); FR= French; DE=German; IT=Italian; PT=Portuguese; ES=Spanish; JP=Japanese; PL=Polish; RU=Russian
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Firm received reports of combustion within the blender system of the Inspiration Ventilator.
FDA root cause
Other
Action
Customer shall be notified by an Urgent Product Safety Notfication letter dated August 12, 2005. The letter shall advise the customer with the specifics of this problem, summarise the fact that the hazard potential in negligible and will advise the customer regarding options for replacement of the Blender Module assembly (part number F710400) and the Air/O2 Filters (part number F910300). The letter will also provide the customers with a set of actions which can be completed immediately in order to ensure the integrity of their gas supplies until their unit has been upgraded.
Quantity in commerce
1,602 units
Distribution
Nationwide and Canada, Europe, Japan, Latin America, Asia/Pacific, Middle East
Product codes and lots
All serial numbers for Models F7100000-XX, F7200000-XX, and F7300000-XX