Z-0023-06
Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface...
Zimmer GmbH / initiated 2005-08-23 / Terminated
Zimmer Inc. began a recall of Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface, small, sterile, for cemented use only in the USA... on . The reason FDA records is "The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0023-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 40 mm dia. articular surface, small, sterile, for cemented use only in the USA, Cat. no. 00430604000.
- Product code
- KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
- Reason for recall
- The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.
- FDA root cause
- Other
- Action
- Consignees were notified via letter dated August 23, 2005 or by email.
- Distribution
- Nationwide, Australia, Canada, England, Germany and Italy.
- Product codes and lots
- All lots.