Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0027-06

Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface...

Zimmer GmbH / initiated 2005-08-23 / Terminated

Zimmer Inc. began a recall of Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface, med, sterile, for cemented use only in the USA... on . The reason FDA records is "The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0027-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface, med, sterile, for cemented use only in the USA, Cat. no. 00430604652.
Product code
KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for recall
The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.
FDA root cause
Other
Action
Consignees were notified via letter dated August 23, 2005 or by email.
Distribution
Nationwide, Australia, Canada, England, Germany and Italy.
Product codes and lots
All lots.