Z-0858-06
GE Signa Advantage nuclear magnetic resonance imaging system
GE Healthcare / initiated 2005-08-19 / Terminated
GE Medical Systems, LLC began a recall of GE Signa Advantage nuclear magnetic resonance imaging system on . The reason FDA records is "During use of GE Signa nuclear magnetic resonance imaging system, a flexible Teflon tube that was part of the magnet vent system ruptured and allowed the magnet to vent into the scan room. Because of helium pressure...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0858-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Signa Advantage nuclear magnetic resonance imaging system.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- During use of GE Signa nuclear magnetic resonance imaging system, a flexible Teflon tube that was part of the magnet vent system ruptured and allowed the magnet to vent into the scan room. Because of helium pressure inside the room, the scan room door could not be reopened after the patient was removed from the scan room.
- FDA root cause
- Other
- Action
- GE Healthcare is issuing an Urgent Safety Notification to all affected customer sites beginning 09/01/05. The notification describes the concurrent condition that must exist in order for the event to occur. Users are warned to be present at all times when scanning patients and to take emergency measures to vent the scan room should the magent quench. Beginning September 2005, all affected customers'' devices will have the Teflon venting tube replaced with a nonflexible venting tube.
- Quantity in commerce
- 644 (516 US, 128 OUS)
- Distribution
- Nationwide and to countries OUS including: United Arab Emirates, Argentina, Australia, Brazil, Canada, China, Germany, Denmark, Egypt, Spain, France, United Kingdom, Ireland, Italy, Jordan, Japan, Korea, Kuwait, Lebanon, Luxembourg, Mexico, Netherlands, Saudi Arabia, Sweden, Turkey and Tiawan
- Product codes and lots
- All GE Signa Advantage MRI magnets with the model numbers 46-260805G1, 46-260805G2, 46-281380G1, 46-281380G2, 46-281380G3, 46-281380G4, 46-281380G5, 46-281380G6, 46-281380G7 and 46-281380G8