Enpath Steerable Sheath, model 10775-003, Sterile EO. Single Use Only. Enpath Medical, Inc., 15301 Highway 55...
Enpath Medical, Inc. began a recall of Enpath Steerable Sheath, model 10775-003, Sterile EO. Single Use Only. Enpath Medical, Inc., 15301 Highway 55 West, Minneapolis, MN 55447, USA. The Enpath... on . The reason FDA records is "Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1573-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Greatbatch Medical
- Product
- Enpath Steerable Sheath, model 10775-003, Sterile EO. Single Use Only. Enpath Medical, Inc., 15301 Highway 55 West, Minneapolis, MN 55447, USA. The Enpath Deflectable Catheter is a felexible tipped percutaneous catheter designed for gaining access to the peripheral and coronary systems. The device features adjustable tip geometry through use of a sliding mechanism with locking feature to deflect the catheter. There are two versions of the tool: one with a luer fitting for Tuohy-Borst valve connection, the other with a hemostasis valve and a sideport infusion line. The kit includes a flexible catheter dilator to facilitate deflectable catheter passage.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Enpath has become aware that some of the Enpath Medical 8 Fr Steerable sheath devices may have non-conforming hemostasis valves in that the valve can leak or the valve was observed to have a small hole.
- FDA root cause
- Other
- Action
- Clinical locations were sent an Urgent Medical Device Recall letter dated 09/02/05 from Enpath Medical, Inc. The letter states the issue, affected lots to be segregated and to discontinue use of. A Bard Electrophysiology Territory Manager will retrieve affected product.
- Quantity in commerce
- 140
- Distribution
- MA
- Product codes and lots
- W34968 and W35440