Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5...
Ekos Corp began a recall of Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per... on . The reason FDA records is "During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0111-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- EKOS Corporation
- Product
- Lysus Infusion System - Lysus Drug Delivery Catheter: The Lysus Infusion System employs high frequency (2-2.5 MHz), low power (0.45 watt average power per transducer) ultrasound to facilitate the delivery of physician-specified fluids in the peripheral vasculature. (This system is also called the UltraFlo and EKOS Peripheral Infusion System in internal documents.) The Lysus Infusion System consists of three main components, a single use Lysus Infusion Catheter System which is comprised of a Drug Delivery Catheter (DDC) and removable Ultrasound Core (USC), and a reusable EKOS PT-3 Control System. During use, fluids are delivered through the side-holes of the DDC. The USC is placed through the central lumen of the DDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS PT-3 Control Unit provides power to the system and the user interface for operator control. A reusable, non-sterile Catheter Interface Cable with Pod connects the EKOS PT-3 Control Unit to the Lysus Infusion Catheter System.
- Product code
- KRA Catheter, Continuous Flush
- Reason for recall
- During two separate procedures, the user experienced difficulty placing the EKOS ultrasound core (USC) into the EKOS drug delivery catheter (DDC) when transiting tight bend radii; specifically the aortic bifurcation. In one case, the USC fractured within the DDC when the user attempted to withdraw the USC after resistance was encountered
- FDA root cause
- Other
- Action
- On 11/22/04 the firm issued letters to those customers that had product. The letters advised of the problem and provided amended use instructions.The firm continued to ship product and on 3/10/05 the firm visited consignees, replacing the catheters.
- Quantity in commerce
- 134 kits
- Distribution
- Kits were distributed to 6 hospitals in the US, 1 each located in AZ, OH, WA, FL, IL, and CA. The firm also distributed to a customer in Germany.
- Product codes and lots
- Serial #s 040914006-004, 040914006-008, 040914007-006, 040914008-003, 040928003-002, 040928003-005, 040928003-009, 040928004-002, 040928005-004, 040928006-002, 040928006-003, 040928006-006, 041013005-004, 040702005-007, 040702005-008, 040719001-005, 040719001-006, 040913002-001, 040913002-002, 040913002-003, 040914007-002, 040914010-001, 040928004-005, 041013006-014, 040928003-012, 040928003-013, 041013005-006, 041013005-007, 041013006-011, 041013006-017, 040719002-003, 040322001-003, 040322001-006, 040322001-013, 040420001-001, 040420001-002, 040420001-004, 040420001-005, 040420001-006, 040420001-007, 040420001-008, 040420001-009, 040420001-010, 040420001-011, 040702005-001, 040702005-002, 040702006-001, 040702006-002, 040928004-004, 040928005-002, 040928005-005, 040928006-004, 040928006-005, 040702005-003, 040702005-004, 040702006-003, 040702006-004, 040719001-001, 040719001-002, 040719002-002, 040719003-001, 040719003-002, 040914007-005, 040914008-002, 040914009-003, 040928004-001, 040928006-009, 040928006-010, 040928007-002, 040702005-005, 040702005-006, 040702006-005, 040719001-003, 040719001-004, 040909003-001, 040909003-002, 040913002-004, 040913002-005, 040914006-007, 040914007-003, 040914007-004, 040914009-004, 040928003-001, 040928004-008, 040928005-006, 040928006-001, 040928007-001, 041013005-005, 041013006-012, 041013006-013, 041119002-001, 041119003-001, 040719001-007, 040914006-003, 040914007-001, 040914008-001, 040914009-001, 040928003-010, 040928004-006, 040928004-007, 041013004-001, 041013004-002, 041013004-003, 041013006-008, 041013006-015, 040420001-003, 040719001-008, 040719002-001, 040914006-001, 040914006-005, 040914006-006, 040914006-009, 040914006-010, 040914008-004, 040928003-003, 040928003-004, 040928003-011, 040928003-014, 040928004-003, 040928005-007, 040928005-008, 041013006-009, 041013006-016, 041013006-018, 050204002-001, 050204002-002, 050204002-003, 050204002-004, 050204002-005, 050209002-001, 050209002-002, 050209002-003, 050209002-004, 050209002-005 Catalog # & Part #: 500-52112, 3383-002; 500-52124, 3393-004; 500-52124, 3393-006; 500-52150, 3393-007