Z-0006-06
Nichols Advantage Invasive Trophoblast Antigen (ITA)Assay Cartridges (Catalog #62-7052)
Nichols Institute Diagnostics / initiated 2005-08-31 / Terminated
Nichols Institute Diagnostics began a recall of Nichols Advantage Invasive Trophoblast Antigen (ITA)Assay Cartridges (Catalog #62-7052) on . The reason FDA records is "Data from internal testing indicates that the Nichols Advantage Invasive Trophoblast Antigen Assay Cartridges (ITA)(Lot No 62-500261) and Hyperglycosylated Human Chorionic Gonadotropin Assay Cartridges (H-hCG)(Lot No...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0006-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nichols Institute Diagnostics
- Product
- Nichols Advantage Invasive Trophoblast Antigen (ITA)Assay Cartridges (Catalog #62-7052)
- Product code
- DHA System, Test, Human Chorionic Gonadotropin
- Reason for recall
- Data from internal testing indicates that the Nichols Advantage Invasive Trophoblast Antigen Assay Cartridges (ITA)(Lot No 62-500261) and Hyperglycosylated Human Chorionic Gonadotropin Assay Cartridges (H-hCG)(Lot No 62-500262) does not maintain its stability over the 12 month expiration dating of the product, and that stability is assured only through 6 months.
- FDA root cause
- Other
- Action
- Withdrawal letter was sent 08/30/2005 which request customers to stop using the product immediately and discard any remaining materials. A response card was included.
- Quantity in commerce
- 22
- Distribution
- Nationwide and China, Germany, & Australia
- Product codes and lots
- Lot#62-500261