Z-0038-06

Datex-Ohmeda S/5 Aespire Anesthesia System 7900

GE Healthcare / initiated 2005-09-27 / Terminated

Datex-Ohmeda Inc began a recall of Datex-Ohmeda S/5 Aespire Anesthesia System 7900 on . The reason FDA records is "Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0038-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Datex-Ohmeda S/5 Aespire Anesthesia System 7900
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
Two conditions may occur with the S/5 Aespire Anesthesia System. The internal auxiliary flowmeter tubing may become disconnected if the auxiliary oxygen supply tubing is occluded and the patient may not receive desired supplemental oxygen which could lead to serious injury if undetected. Also the O2 switch may leak or fail.
FDA root cause
Other
Action
Urgent Medical Device Correction letter dated 09/23/05 was sent to all customers. The letter describes the issues, affected device serial numbers and recommends that customers DO NOT USE the internal auxiliary oxygen flowmeter, if installed in their system, and to use alternate auxiliary oxygen sources, i.e. wall-mounted oxygen flow meter, in the interim. A GE service representative will contact the customers to schedule an upgrade, performed at no charge.
Quantity in commerce
16
Distribution
CA, CO, FL, GA, IL, IN, MI, MS, NM, OK, PA, PR, SD, UT & WA. OUS to: Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Canada, Caribbean, Czech Republic, Chile, China, Colombia, Dominican Republic, Ecuador, Finland, France, Germany, Ghana, Great Britain, Hong Kong, India, Indonesia, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Mexico, Oman, Palestine, Peru, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, United Arab Emerites, Venezuela & Vietnam.
Product codes and lots
ANCJ00100 thru ANCJ00113, ANCJ00116, ANCJ00118