Z-0102-06

Accumetrics part number 29001 VerifyNow Instrument, printe power supply

Accumetrics, Inc. / initiated 2005-09-13 / Terminated

Accumetrics Inc began a recall of Accumetrics part number 29001 VerifyNow Instrument, printe power supply on . The reason FDA records is "In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0102-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accumetrics, Inc.
Product
Accumetrics part number 29001 VerifyNow Instrument, printe power supply
Product code
JOZ System, Automated Platelet Aggregation
Reason for recall
In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply.
FDA root cause
Other
Action
Letters and replacement power supplies were sent to consignees by Fedex, 2-day with proof of delivery. End-user customers were instructed to return the affected power supply to Accumetrics, using the provided pre-paid shipping carton. Consignees who do not return the power supply within two weeks of the notification will be contacted by phone to assess compliance with the recall and to determine the status of the product at the site.
Quantity in commerce
11
Distribution
CA, FL, IL, LA, NH, NV, NY, OK, WA
Product codes and lots
lot number G-13850