Z-0891-06

CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199). Cardiac Resynchronization Therapy Defibrillator...

Boston Scientific Corporation / initiated 2005-09-30 / Terminated / root cause: supplier

Guidant Corporation began a recall of CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant... on . The reason FDA records is "Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0891-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
CONTAK RENEWAL 4 CRT-D (model no. H190, H195, H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA. CONTAK RENEWAL 4 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
Devices manufactured with batteries built on April 26 or 27, 2005, by a single operator, may have battery welds that may have been performed at settings outside of typical manufacturing specifications at the supplier.
Root cause tag
supplier
FDA root cause
Other
Action
Guidant issued a Physicain letter, dated 10/25/05 to implanting/following physicians. The letter states the issues and reommends patient follow-up care every three months, as stated in published labeling. An Advisory Update, dated 12/20/05, was sent to physicians that describes the issue, states that here have been zero reports of battery weld failures on implanted devices and the results of laboratory testing (accelerated testing) in which they tried to induce the failure. The results of the testing induced 2 battery weld post failures. A projected rate of occurence was also given. Patient management recommendations have not changed.
Quantity in commerce
70
Distribution
Nationwide (AL, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI). OUS to include: Australia, France, Germany, India, Ireland, Israel, Italy, Netherlands, Norway, South Africa, Spain, Switzerland, Turkey, United Kingdom.
Product codes and lots
Model H190, serial numbers: 310258, 310260, 310293, 310312, 310317, 310319, 310320, 310321, 310322, 310343,  310353, 310355, 310381, 310383, 310386, 310387, 310388,  310389, 310390.   Model H195, serial numbers: 350236, 350239, 350245,  350247, 350248, 350250, 350252, 350255, 350261, 350264,  350276, 350279, 350280, 350282, 350284, 350319, 350330,  350331, 350332, 350333, 350334, 350336, 350342, 350344,  350350, 350351, 350353, 350354, 350355, 350360, 350362,  350363, 350364, 350365, 350403, 350407, 350442, 350451,  350480, 350484, 350503, 350570, 350649.  Model H199, serial numbers: 320330, 320331, 320361, 320366, 320370, 320415, 320417, 320445.