Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor...
Datascope Corp began a recall of Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological... on . The reason FDA records is "Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0176-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch color display and has a standard configuration of a 3 or 5 lead ECG, Masimo SET SpO2, non-invasive blood pressure, respiration, continuous temperature and IV Drug Calculations. Optional digital displays are provided for invasive blood pressure (up to four) anesthetic agents, O2 and NO2, ST, and CO2.
- Product code
- DRT Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)
- Reason for recall
- Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
- Quantity in commerce
- 2318 monitors
- Distribution
- The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
- Product codes and lots
- All Spectrum Monitors that have the following Software Versions: A.19, A.20, B.23, B.24, B.25, B.27, C.18, D.07, E.07, F.09, F.16, F.17, F.26.