Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0177-06

Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor...

Datascope Corp. / initiated 2005-09-07 / Terminated / root cause: software

Datascope Corp began a recall of Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following... on . The reason FDA records is "Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0177-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Passport 2 with View 12 ECG Analysis Module. It is a transportable, multi-parameter physiological monitor designed to monitor and display the following physiological parameters: ECG, Heart Rate derived from selected sources (Sp)2, ECG, IBP and NIBP), SpO2 level, ST Segment, Arrhythmia, Blood pressure (both invasive and non-invasive), Respiration rate (dirived from ECG or CO2), inspired or expired CO2, temperature , and gases (i.e. five anesthetsia gases, O2, N2O, and CO2)
Product code
DRT Monitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)
Reason for recall
Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.
Root cause tag
software
FDA root cause
Other
Action
Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.
Quantity in commerce
979 monitors
Distribution
The monitors have been distributed domestically and internationally to hospitals and clinics. There are 10 Passport 2 monitors distributed to Naval Hospital, Camp Pendleton, CA. There are 12 Spectrum monitors distributed to two US Govt facilities: VA Medical Center, Hines, IL and VA Medical Center, Syracuse, NY.
Product codes and lots
Passport 2 Monitors with software versions: S.10, T.15, T.31, V.07, W.09, W.13, W.14.