Z-1549-2013Class II

Aperio ScanScope System. Aperio ScanScope Models, XT, AT, AT Turbo, CS and CSO. Automated digital slide...

Leica Biosystems Imaging, Inc. / initiated 2013-02-22 / Terminated

Aperio Technologies Inc began a recall of Aperio ScanScope System. Aperio ScanScope Models, XT, AT, AT Turbo, CS and CSO. Automated digital slide creation and viewing device. It is intended to create... on . FDA classified it as Class II and gives the reason as "Aperio initiated this recall because the User Guides incorrectly state the devices had been cleared for specific applications. The User Guides should state the devices are for Research Use Only and are not for use in...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1549-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Leica Biosystems Imaging, Inc.
Product
Aperio ScanScope System. Aperio ScanScope Models, XT, AT, AT Turbo, CS and CSO. Automated digital slide creation and viewing device. It is intended to create and view electronic image files representative of tissue specimens fixed to glass slides.
Product code
IBM Microscope, Phase Contrast
Reason for recall
Aperio initiated this recall because the User Guides incorrectly state the devices had been cleared for specific applications. The User Guides should state the devices are for Research Use Only and are not for use in diagnostic procedures.
FDA root cause
No Marketing Application
Action
Aperio sent an Urgent Medical Device Recall letter dated April 22, 2013 to all affected consignees via certified USPS mail. The letter included the reason for the recall, recall product information, instructions, acknowledgement form, contact information ( Aperio Technical Support at 1-866-478-3999), and product labeling. The consignees were requested to acknowledge receipt of the Medical Device Correction letter. In addition, an Acknowledgement Form is required to be signed and e-mailed back to the Manufacturer as evidence of letter receipt. For questions regarding this recall call 760-304-6211, ext 6211.
Quantity in commerce
364
Distribution
Nationwide Distribution - including all states, none internationally .
Product codes and lots
The following serial numbers of the Aperio ScanScope systems are affected in the USA by this recall:  ¿ScanScope XT ScanScope AT ScanScope ATT ScanScope es, esa  1223, 1273, 1278, 1277, 1282, 1593,1605, 1648,1657,1660, 5064,5068,5108,5155,5170,5171  1285,1286,1287,1293,1296, 1619, 1620, 1662,1663,1666, 5176,5181,5187,5196,5198,5202,  1299,1300,1301,1302,1304, 1623, 1633, 1671, 1672, 1673, 5204,5205,5233,5236,5237,5241,  1311,1312,1314,1315,1316, 1634,1635, 1676, 1678, 1679, 5244,5245,5246,5248,5253,5262,  1318,1319,1320,1321,1322, 1636, 1637, 1690, 1692, 1695, 5266,5267,5269,5270,5272,5276,  1324,1325,1327,1328,1330, 1638,1640, 1698,1700,1718 5277,5279,5283,5287,5288,5293,  1332, 1333, 1335, 1336, 1338, 1643,1645, 5294, 5295, 5302, 5307, 5308, 5309,  1339, 1340, 1342, 1343, 1345, 1649,1654, 5314,5315,5316,5317,5324,5327,  1350,1352,1357,1358,1359, 1656,1664, 5328,5329,5330,5331,5333,5334,  1366, 1368, 1369, 1370, 1372, 1686, 1687, 5336,5339,5340,5341,5345,5347,  1382,1383,1392,1394,1396, 1688, 1689, 5348,5352,5353,5356,5367,5371,  1397, 1399, 1400, 1402, 1404, 1696, 1699, 5374,5379,5381,5388,5390,5394,  1405,1407,1409,1410,1413, 1703,1705, 5396,5402,5403,5404,5405,5406,  1417,1418,1419,1420,1421, 1707, 1708, 5408,5409,5415,5416,5418,5419,  1422, 1423, 1424, 1425, 1426, 1709,1711, 5433,5440,5452,5456,5457,5459,  1436,1441,1442,1444,1445, 1714,1716, 5462,5465,5466,5467,5468,5469,  1446, 1448, 1449, 1450, 1452, 1719,1721, 5470,5473,5475,5476,5477,5478,  1453, 1454, 1456, 1457, 1458, 1723,1724, 5482,5485,5486,5489,5490,5492,  1460,1461,1466,1475,1476, 1725,1727, 5493,5504,5515,5516,5517,5518,  1481,1482,1486,1487,1490, 1728,1729, 5520, 5529, 5537, 5540, 5552, 5566,  1492, 1499, 1500, 1502, 1503, 1731,1733, 5567,5574,5578,5581,5587,5590,  1505,1506,1507,1508,1509, 1735, 1736, 5591,5592,5593  1511,1512,1514,1515,1516, 1739,1744  1517,1519,1520,1522,1523,  1525, 1527, 1528, 1532, 1534,  1535, 1536, 1537, 1538, 1540,  1542,1543,1544,1545,1546,  1547,1548,1549,1550,1551,  1552, 1553, 1554, 1556, 1557,  1559,1560,1563,1564,1569,  1570, 1571, 1575, 1576, 1582,  1586,1587,1588,1590,1591,  1596,1597,1611,1627

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1549-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1549-2013Class IITerminatedVoluntary: Firm initiated