Z-0303-06
3G Storm Series Power Wheelchair with Power Tilt Option
Invacare GmbH / initiated 2005-10-11 / Terminated
Invacare Corporation began a recall of 3G Storm Series Power Wheelchair with Power Tilt Option on . The reason FDA records is "When the powered seating system is tilted completely in the rearward position, a potential pinch point is created between the seat and the seat frame". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0303-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invacare GmbH
- Product
- 3G Storm Series Power Wheelchair with Power Tilt Option.
- Product code
- ITI Wheelchair, Powered
- Reason for recall
- When the powered seating system is tilted completely in the rearward position, a potential pinch point is created between the seat and the seat frame.
- FDA root cause
- Other
- Action
- The recalling firm sent letters to all consignees.
- Quantity in commerce
- 16 devices
- Distribution
- The product was sent to distributors located in AL, GA, CA, IL, KS, MN, NC, PA, TX, and VA.
- Product codes and lots
- Serial numbers: 05G066858, 05G066859, 05G071801, 05G071803, 05G077773, 05G079666, 05G089011, 05G089012, 05G092274, 05G105434, 05G106984, 05G111332, 05G117359, 05G121857, 05G125563, 05H144898