Jet Medical 4F Single Lumen PICC Kit. Catalog number JP7-4

Medical Components, Inc. / initiated 2005-09-15 / Terminated / root cause: labeling and UDI

Medical Components, Inc dba MedComp began a recall of Jet Medical 4F Single Lumen PICC Kit. Catalog number JP7-4 on . The reason FDA records is "The product is labelled as containing a 4F catheter when the trays/kits actually contain a 5F catheter. The catheters will not fit into the sheath-dilator that was included in the tray. mislabeled as 4F". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0104-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medical Components, Inc.
Product
Jet Medical 4F Single Lumen PICC Kit. Catalog number JP7-4
Product code
LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
Reason for recall
The product is labelled as containing a 4F catheter when the trays/kits actually contain a 5F catheter. The catheters will not fit into the sheath-dilator that was included in the tray. mislabeled as 4F
Root cause tag
labeling and UDI
FDA root cause
Other
Action
On 9/15/05, the recalling firm telephoned the two distributors to inform them of the problem and the need to return the product.
Quantity in commerce
88 units
Distribution
The products were shipped to distributors in OH and MO.
Product codes and lots
Lot number MABY950