Z-0304-06
Dade Actin FS Activated PTT Reagent. Reagent and Control Partial Thromboplastin Time. Catalog number B4218-100
Siemens Healthcare Diagnostics Inc. / initiated 2005-09-23 / Terminated
Dade Behring , Inc. began a recall of Dade Actin FS Activated PTT Reagent. Reagent and Control Partial Thromboplastin Time. Catalog number B4218-100 on . The reason FDA records is "prolonged aPTT results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0304-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dade Actin FS Activated PTT Reagent. Reagent and Control Partial Thromboplastin Time. Catalog number B4218-100
- Product code
- GGW Test, Time, Partial Thromboplastin
- Reason for recall
- prolonged aPTT results
- FDA root cause
- Other
- Action
- The recalling firm issued a letter dated 9/30/05 informing the customers of the problem and instructing them to discontinue use of the product.
- Quantity in commerce
- 4176 packs
- Distribution
- The products were shipped to medical facilities and/or laboratories nationwide.
- Product codes and lots
- Lot number 527263 exp 2/3/06