Z-0466-06

Omron brand 3-Way Instant Thermometer, Model Numbers MC-600 (USA) and MC-600CAN (Canada); an OTC device used...

Omron Healthcare Co., Ltd. / initiated 2005-10-25 / Terminated

Omron Healthcare, Inc. began a recall of Omron brand 3-Way Instant Thermometer, Model Numbers MC-600 (USA) and MC-600CAN (Canada); an OTC device used to take an accurate patient temperature... on . The reason FDA records is "Omron is recalling the Omron 3-Way Instant Thermometers due to possible overheating of the probe tip". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0466-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Omron Healthcare Co., Ltd.
Product
Omron brand 3-Way Instant Thermometer, Model Numbers MC-600 (USA) and MC-600CAN (Canada); an OTC device used to take an accurate patient temperature measurement in 4-6 seconds typically in the oral, rectal or under arm modes when the ambient temperature is between 50 degrees F and 93.2 degrees F; Distributed by: Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061, Made in Philippines; UPC 73796-43600
Product code
FLL Continuous Measurement Thermometer
Reason for recall
Omron is recalling the Omron 3-Way Instant Thermometers due to possible overheating of the probe tip.
FDA root cause
Other
Action
Omron sent recall letters dated 10/21/05 to all of their direct accounts, informing them that the probe tip could overheat and possibly burn the person whose temperature is being taken. The accounts were instructed to check the lot number on their thermometers immediately to see if they fall within the range of affected thermometers (lots beginning with 01-32, 01-36, 01-37 and 01-38). The accounts were instructed to call Omron at 1-800-634-4350 if their unit is affected to get instructions on how to return the unit to Omron for a refund. Omron issued a press release on 10/24/05 to alert consumers who may have purchased the thermometer at a retail pharmacy, and has posted both the press release and the recall letter on their website at www.omronhealthcare.com.
Quantity in commerce
Approximately 48,000 units
Distribution
Nationwide and internationally to Canada, Chile, Venezuela and Singapore
Product codes and lots
Model Numbers MC-600 (USA) and MC-600CAN (Canada) Lot numbers beginning with 1-32, 1-36, 1-37, 1-38.