Z-0089-06
Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545...
Medrad, Inc. / initiated 2005-10-07 / Terminated
Medrad Inc began a recall of Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550... on . The reason FDA records is "overhead system can fall due to the failure of a retaining ring". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0089-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medrad, Inc.
- Product
- Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- overhead system can fall due to the failure of a retaining ring.
- FDA root cause
- Other
- Action
- The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/7/05 to all customers. The notice informs customers of the problem and cautions anyone to be under the system when in use. The notice also informs the customer that a firm representative will schedule a time to perform the upgrade. On 10/13/05, the recalling firm also issued a notification to scanner manufacturers alerting them of the problem.
- Quantity in commerce
- 2080 units
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Austria, Belgium, Chile, Czech, Denmark, Spain, Finland, France, Great Britain, Ireland, Israel, Italy, Netherlands, Norway, Sweden, Austrailia, Kuwait, Saudia Arabia, Canada, China, Hong Kong, India, Korea, Singapore, Taiwan, Brazil, Columbia, Cuba, Mexico, and Japan.
- Product codes and lots
- All serial numbers