Z-0089-06

Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545...

Medrad, Inc. / initiated 2005-10-07 / Terminated

Medrad Inc began a recall of Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550... on . The reason FDA records is "overhead system can fall due to the failure of a retaining ring". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0089-06
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medrad, Inc.
Product
Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
overhead system can fall due to the failure of a retaining ring.
FDA root cause
Other
Action
The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/7/05 to all customers. The notice informs customers of the problem and cautions anyone to be under the system when in use. The notice also informs the customer that a firm representative will schedule a time to perform the upgrade. On 10/13/05, the recalling firm also issued a notification to scanner manufacturers alerting them of the problem.
Quantity in commerce
2080 units
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Austria, Belgium, Chile, Czech, Denmark, Spain, Finland, France, Great Britain, Ireland, Israel, Italy, Netherlands, Norway, Sweden, Austrailia, Kuwait, Saudia Arabia, Canada, China, Hong Kong, India, Korea, Singapore, Taiwan, Brazil, Columbia, Cuba, Mexico, and Japan.
Product codes and lots
All serial numbers