Z-0416-06
Vysis LSI p16 (9p21)/CEP 9 (9p11-q11) Dual Color Probe Set; a locus specific identifier DNA probe consisting...
Abbott Molecular, Inc. / initiated 2005-10-28 / Terminated / root cause: labeling and UDI
Abbott Molecular began a recall of Vysis LSI p16 (9p21)/CEP 9 (9p11-q11) Dual Color Probe Set; a locus specific identifier DNA probe consisting of a mixture of the LSI p16 probe labeled with a... on . The reason FDA records is "The DNA-Probe has a target size of ~300 Kb region instead of the labeled ~190 Kb region". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0416-06
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Molecular, Inc.
- Product
- Vysis LSI p16 (9p21)/CEP 9 (9p11-q11) Dual Color Probe Set; a locus specific identifier DNA probe consisting of a mixture of the LSI p16 probe labeled with a SpectrumOrange and the CEP 9 probe labeled with a SpectrumGreen fluorophore, accompanied with LSI/WCP Hybridization Buffer; 20 evaluations; Vysis Inc., Downers Grove, IL 60515; order number 32-190078
- Product code
- MAO Dna-Probe, Human Chromosome
- Reason for recall
- The DNA-Probe has a target size of ~300 Kb region instead of the labeled ~190 Kb region.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Abbott Molecular sent letters dated October 2005 to their direct laboratory accounts on 10/28/05 via first class certified mail. The accounts were informed of the discrepancy between the actual size and region of coverage produced by the affected lots and that described in the product literature. The accounts were instructed to assess the impact of this discrepancy on the use of the product in their laboratory and to retain a copy of the letter with any of the affected product remaining in their inventory.
- Quantity in commerce
- 749 kits
- Distribution
- Nationwide, Germany, Japan, Canada, Korea, India, Malaysia, Australia, Hong Kong, Chile & Singapore
- Product codes and lots
- order number 32-190078, lots 42919, 45942, 53417, 57026, 59938, 60858
FIRST ARTICLE / WEEKLY